Cardiac and vascular prostheses

Telechargé par drfrangi
Cardiac and vascular
prostheses
GRAEME L. HAMMOND
The development of vascular protheses in the early 1950s and of cardiac prostheses in the
early 1960s extended surgical treatment to diseases that had been previously beyond this field
and set the stage for the development of the artificial assist devices that are now at the
forefront of surgical research. Although room for improvements in materials and design
remains, the clinical use of these devices has reduced the mortality associated with ruptured
aortic aneurysm from 100 per cent to 30 per cent, and that of symptomatic aortic stenosis from
100 per cent to 5 per cent.
VASCULAR PROTHESIS
In general, grafts of lower porosity are stiffer, while higher porosity grafts are softer. These are
important considerations when treating patients receiving heparin or when operating on
friable or extensively diseased vessels. The use of the wrong graft in any particular situation
can cause life-threatening complications. The two most commonly used prosthetic materials
are Dacron and polytetrafluoroethylene (Teflon). Dacron grafts come in knitted and woven
varieties: the former are of high porosity and are easier to sew, but leak when the vascular
clamps are removed or until clot forms in the graft interstices. Woven grafts are of low
porosity, and more difficult to sew but bleeding through the graft is minimized or eliminated.
Teflon grafts are not porous and do not leak but tend to be less compliant and are therefore
more difficult to sew than Dacron.
It was originally thought that a high porosity (knitted) graft was necessary to allow endothelial
cells and their subendothelial matrix to invade and adhere to the inner graft surface, reducing
thromboembolism and increasing patency. Although prosthetic vascular grafts in canine and
non-human primate models may develop a neoendothelium, re-endothelialization occurs only
sporadically or not at all in man. The patency of woven and knitted grafts was compared in a
study in which bifurcation grafts, one limb of which was knitted and one limb woven, were
implanted in 143 consecutive patients with aortoiliac atherosclerosis or aneurysms. There was
no difference in patency rates between the woven and knitted limbs during observation
periods of 1 month to 2 years.
Knitted grafts are designed for abdominal and peripheral vascular procedures. In general,
they are easily sewn to blood vessels because of their soft, compliant nature. The compliance
of the graft allows a snug fit, even when the native vessel is heavily calcified, and minimizes
bleeding at anastomotic sites. However, removal of the clamps is followed by bleeding through
the graft until clot forms in the interstices. Blood loss can be considerable, and use of these
grafts is contraindicated in patients receiving heparin, due to the likelihood of uncontrollable
haemorrhage. A modification of the standard knitted graft is the double velour graft (Medox)
which incorporates a velour inner and outer pile to enhance the incorporation of clot in the
graft interstices. Dacron grafts of less than 5 mm in diameter should not be used since they are
associated with a high incidence of thrombosis. Such grafts are therefore not suitable for the
treatment of obstructive disease below the knee.
Woven grafts are designed specifically for use in patients who require systemic heparinization
during graft insertion. They are thus ideally suited for replacement of the thoracic aorta. Graft
stiffness is not usually a problem when reconstructing the aorta after resection of fusiform or
sacular aneurysms, as the resection margins tend to be thick and fibrous, and incorporate
sutures and graft well. In patients with ascending thoracic aortic dissection, however, the
aortic tissues are extremely thin, tenuous, and friable. Attempts to sew a stiff, non-compliant
graft into these tissues are often fraught with further tearing, sometimes to the point of
avulsion of a suture line. On the other hand, low porosity is a requirement since patients
undergoing surgery for Type A dissections are on cardiopulmonary bypass and often develop
severe pre- or intraoperative coagulopathies. A satisfactory compromise has been developed
by Medox in the form of the low porosity Veri-Soft woven graft. In this case, the Dacron yarn is
soft enough to allow good flexibility, but the weave is tight enough to prevent haemorrhage
through the interstices. The Veri-Soft graft is effective as a replacement for the ascending
aorta; however, we also toughen the aorta with glutaraldehyde to ensure the stability of the
aortic tissues prior to performing the anastomosis.
All grafts, whether knitted or woven, can be preclotted, thereby eliminating the problem of
haemorrhage through the graft. Although routinely preclotting of Veri-Soft woven grafts is not
required before their use in the thoracic aorta, this may be necessary when a thoracic
dissection is superimposed upon coagulopathic states such as chronic aspirin therapy. If only
heparinized blood is available, the graft can be soaked in thawed fresh frozen plasma and then
autoclaved. Although this results in some stiffening, it guarantees a water-tight graft.
The smooth, non-adherent, non-wettable surface of Teflon theoretically does not allow
platelets and fibrinous material to adhere; grafts of smaller diameter can therefore be placed
in peripheral vessels. An additional theoretical advantage of Teflon is that it is non-porous to
blood elements, and bleeding through the graft does not occur. However, Teflon grafts are non-
compliant, and bleeding through suture holes and gaps between sutures is common. Bleeding
through suture holes can be minimized by using a smaller diameter needle and Gore-Tex
suture material that expands when wet. While Dacron grafts are pleated, Teflon grafts are not,
and they tend to kink more easily. However, some Teflon grafts are now being manufactured
with external ribbing to help prevent kinking.
The superiority of Teflon over Dacron regarding long-term patency is controversial. There is
no difference in the development of intimal hyperplasia at anastomotic sites or in the patency
of aortic bifurcation grafts made from the two materials. However, there appears to be a
somewhat higher patency rate for Teflon grafts when used in the axillofemoral position or in
the femoral-popliteal-tibial position. Neither Teflon or Dacron is comparable in patency to
reversed saphenous vein.
New developments
New graft developments include the intraluminal graft, the composite graft, human umbilical
vein grafts, endothelial cell seeded grafts, and collagen-impregnated grafts.
The intraluminal graft was designed for use in the thoracic aorta and to aid in keeping aortic
cross-clamp times to a minimum. The ends of the graft are firm. The prosthesis is inserted into
the aorta and occlusively ligated in place with ties placed around the aorta at the proximal and
distal ends of the graft. It is always more difficult than it first appears to insert the graft into
the aorta, and a graft substantially smaller in diameter than the aorta should be used. Severe
atherosclerosis or tree barking associated with luetic aortitis makes it difficult to obtain a tight
fit between the graft ends and the aorta, and leaking often occurs through the areas of ligation.
In patients with thoracic dissection,tying down the ties can tear the aorta.
Composite grafts are woven Dacron grafts that are sewn to either a porcine or mechanical
valve. These are used for operations on the right ventricular outflow tract in children or to the
ascending aorta in adults. In children it is common to use a porcine valve conduit to minimize
thromboembolism, although the incidence of degradation of these valves in children is high. In
adults, a composite graft using a mechanical valve (usually a St Jude valve) is always
recommended to eliminate the problem of valve degradation and the associated difficulty of
re-replacement. The St Jude valve conduit must be autoclaved before use.
Glutaraldehyde-stabilized human umbilical vein has been used as a conduit for grafting small,
peripheral vessels. In a study of 218 patients undergoing lower limb revascularization, the 3-
year patency rate of human umbilical vein was comparable to that of polytetrafluoroethylene
and considerably lower than that of saphenous vein. Human umbilical vein has, therefore
been not widely used as a bypass conduit for small vessels.
There is a considerable body of literature describing techniques for seeding Dacron or
polytetrafluoroethylene grafts in vitro with endothelial cells. Clinical trials have not been
performed, but results in animals indicate that clot formation is inhibited in such
endothelialized prostheses. At the present time, there are many problems that make this
approach impractical including the time required to endothelialize the graft and rejection
problems which occur if the patient's own endothelial cells are not used. Nevertheless, this
approach may represent an advance in prosthetic graft development for future use in small
vessel reconstruction.
Impregnation of Dacron grafts with collagen eliminates bleeding through interstices and the
need for preclotting; they are only weakly immunogenic.
CARDIAC PROSTHESES
The decision to use any particular heart valve is often based on personal clinical experience.
Although mechanical valves offer satisfactory haemodynamic performance, the associated
risks of thromboembolism have promoted the development of biological valves which, by
partially decreasing the non-biological–biological interface, are less thrombogenic.
Several clinical studies have shown porcine valves are less thrombogenic than are mechanical
valves. However, biological valves themselves are also associated with problems, the most
ominous of which is tissue failure. The risk of failure in the biological valves must therefore be
weighed against the risks of malfunction, thromboembolism, and anticoagulation-related
haemorrhage associated with mechanical valves. Porcine valves are obtained from freshly
slaughtered pigs and are treated so that the final product is a non-living valve, mounted on a
Dacron stent. Upon removal from the pig, the endothelial cells quickly die and are easily
removed by washing. The immunogenic subendothelial elastin layer is removed by treating
the valve with the enzyme ficin. The sole remaining layer is collagen, which has sufficient
species homogeneity to prevent rejection. Since native collagen lacks the strength to withstand
arterial pressures over millions of cycles it is strengthened by cross-linking with
glutaraldehyde. The valve is then sewn to the stent and stored in 0.065 per cent glutaraldehyde
solution.
The Starr Edwards mechanical valve has undergone many modifications since its introduction
in 1961. Its current configuration consists of a silastic ball, stainless steel struts, and a Dacron
sewing ring. The Starr Edwards valve has the longest history of continuous use of any valve,
and it remains a mainstay of cardiac surgery 30 years after its initial introduction. However,
an inherent problem with ball and cage design valves is the high profile of the cage and
tertiary orifice obstruction by the ball.
The Starr Edwards valve has three orifices (Fig. 1) 1658: the primary orifice is the orifice
through which the blood must go when leaving the left ventricle. The secondary orifice is the
orifice subtended by the angle made from the ball in the open position and the primary orifice.
The height of the cage determines the area of the secondary orifice. In a properly designed
valve, the areas of the primary and secondary orifices will be equal. The size of the tertiary
orifice, between the perimeter of the ball in the open position and the walls of the aorta,
cannot be controlled by valve design, and this can cause obstruction to blood flow. Insertion of
a valve with a large primary orifice will produce a small tertiary orifice due to the size of the
ball; conversely, a large tertiary orifice will be associated with a small primary orifice. Because
of the dual problems of tertiary orifice obstruction and high cage profile, the Starr Edwards
valve should not be used in the aortic position in patients with narrow aortic roots, or in the
mitral position in patients with small left ventricular chambers.
To obviate the problems associated with ball and cage design valves, the so-called low profile
valves were developed. The most commonly used of these are the St Jude bileaflet valve, made
from pyrolight carbon with a Dacron sewing ring, and the Bjork-Shiley monoleaflet valve,
which has a pyrolight carbon disc, stainless steel trapping ring, and Dacron sewing ring. There
is no statistically significant difference in long-term clinical outcome of patients treated with
each of the three types of mechanical valves.
A longitudinal analysis of biological and mechanical valves inserted at the Yale-New Haven
Hospital over an 11-year period was unable to demonstrate a clear advantage to the
1 / 8 100%

Cardiac and vascular prostheses

Telechargé par drfrangi
La catégorie de ce document est-elle correcte?
Merci pour votre participation!

Faire une suggestion

Avez-vous trouvé des erreurs dans linterface ou les textes ? Ou savez-vous comment améliorer linterface utilisateur de StudyLib ? Nhésitez pas à envoyer vos suggestions. Cest très important pour nous !