Page | 5
Using 5 MHz simplifies the calibrator circuitry. Also, at this frequency, an oscilloscope can also be
used to set the input voltage from a signal generator fairly accurately.
However, one must note that the AD8307 can have small variations with frequency.
Also, the waveform is very important. The AD8307 is designed to respond to sine wave shapes and is
therefore most accurate with sine wave inputs. However, a square wave can also be used and it has
been found that at an input level of around -20dBm, a square wave of half the peak to peak amplitude
of a sine wave gives the same DC output with a high degree of accuracy. This characteristic behaviour
was documented by Bob Kopskie in a QEX article, Jan/Feb 2004 titled “A Simple RF Power
Calibrator”. See reference 2.
The peak to peak amplitude of a square wave can be measured with a DVM meter. So a square wave
fed into the AD8307 can be used to calibrate the AD8307. The CAL1 calibrator makes use of this
unusual response to a square wave. It should be noted that this relationship only holds around an
input level of -20dBm to -30dBm. The AD640 also exhibits the same behaviour. Fig 23 in the AD640
datasheet shows a graph of the deviation of the square wave to sine wave relationship at other levels.
Because of this, you cannot use a square wave to test the linearity of the AD8307 over its full range of
90dB. The use of a square wave will give misleading results making it look as if the chip is not very
linear. Because of this, the CAL1 calibrator also produces a 5 MHz sine wave which is adjustable.
Used in conjunction with an accurate stepped
attenuator, one can check the linearity of the meter
over a much wider range. The CAL1 sine wave is
generated by passing the square wave through a low
pass filter. Although this is reasonably simple, it does
mean that some harmonic energy can still be part of the
waveform making this only a moderately accurate
method. A good clean sine wave from a signal
generator, of known amplitude, is the best way to test
the linearity of the power meter.
However the CAL1 calibrator is a good compromise
between cost and performance.
It should be noted that the attenuation at UHF is not the same as at VHF in some attenuators. Leakage
across attenuator components can be much higher at UHF due to the intrinsic capacitance of standard
off the shelf resistors. The AQZ attenuators use surface mount resistors but these are still not as good
as the specially designed resistors such as those used in HP professional attenuators. These resistors
have a small shield foil wrapped around the body and small tabs are adjusted in the factory to ensure
non-reactive 50 ohm impedance is maintained over a wide frequency range.
With a high bandwidth RF chip, like the AD8307, an imbalance in the harmonic content of a signal
reaching its input can produce misleading results. For example, an attenuator which produces a 10dB
drop at 5 MHz may not produce a 10 dB drop at, say, the 10th harmonic of 50 MHz. This imbalance
effectively alters the waveform and the AD8307, responding to the higher level 50 MHz harmonic,
may appear not to be as linear as predicted in the data sheet. Unless one has accurate test equipment
and excellent test loads and attenuators, the performance of the AD8307 may seem to be not as good
as the datasheets. This is not a fault of the AD8307 but rather a lack of good test gear capable of
working accurately up to 500 MHz with minimal reactive impedance problems and resonances.
The RFPM1 is a very wideband device so please bear that in mind when calibrating the RFPM1.
Referring to Fig 1, pin 3 of the AD8307 IC has an optional 100nf capacitor (C8) connected to ground.