Telechargé par NAIMA EL HOUET

Rapid LC-IDMS/MS Method for Serum Creatinine Validation

publicité
Clinical Biochemistry 43 (2010) 1158–1162
Contents lists available at ScienceDirect
Clinical Biochemistry
j o u r n a l h o m e p a g e : w w w. e l s e v i e r. c o m / l o c a t e / c l i n b i o c h e m
Development and validation of a rapid liquid chromatography isotope dilution
tandem mass spectrometry (LC-IDMS/MS) method for serum creatinine
Pierre-Olivier Hétu ⁎, Marie-Ève Gingras, Bernard Vinet
Department of Biochemistry, Hôpital Notre-Dame, Centre hospitalier de l'Université de Montréal, Montréal, Québec, Canada
a r t i c l e
i n f o
Article history:
Received 14 January 2010
Received in revised form 17 May 2010
Accepted 29 May 2010
Available online 8 June 2010
Keywords:
Serum creatinine
LC-MS/MS
Isotope dilution
Reference method
a b s t r a c t
Objectives: To develop an isotope dilution liquid chromatography tandem mass spectrometry (LC-IDMS/
MS) method for the standardization of serum creatinine.
Design and methods: Supernatants obtained by protein precipitation were injected into a LC-MS/MS.
Chromatography was performed on a cation exchanger pre-column in normal phase isocratic mode. Creatinine
and creatinine-D3 were quantified using ion transitions of m/z 114→44 and 117→47, respectively.
Results: The method was calibrated with the NIST Standard Reference Material 914a and was found to be
exact in analyzing the certified reference material SRM 967 from NIST (97.1 ± 0.9% and 102.1 ± 0.9% of target
value for levels 1 and 2, respectively) and by calculating recuperation of spiked creatinine in 10 different
patient samples (103.6 ± 4.1%). Intra-assay imprecision was 0.9% at both 64.6 and 354 μmol/L creatinine,
while inter-assay imprecisions were 1.9% and 1.8%. Absence of ion suppression was confirmed by spiking
experiments. The method was shown to be free of carryover. A good correlation was obtained between the LCMS/MS method and a Jaffe method run on an automated analyzer (r = 0.999).
Conclusions: We have developed a fast and simple method for the quantification of serum creatinine by
isotope dilution tandem mass spectrometry (IDMS/MS) and we propose that this method can be used as a
reference method by laboratories that wish to validate their serum creatinine automated assay.
© 2010 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.
Introduction
The prevalence of chronic kidney disease (CKD) is estimated at
approximately 26 million people in the United States [1], and about 2.1
million in Canada [2]. Early identification and treatment of affected
patients is critical because CKD is associated with increased morbidity
and mortality, notably from cardiovascular complications [2–4]. The
glomerular filtration rate (GFR) is the best marker of renal function. GFR
cannot be measured directly, but it can be estimated by the clearance of
inert exogenous compounds that are eliminated exclusively by filtration
through the kidneys, such as inuline or 125I-iothalamate, or more easily
by the clearance of endogenous compounds that are predominantly
cleared by glomerular filtration, such as creatinine [5]. Creatinine is
generated from creatine by normal muscular metabolism at a relatively
constant rate, proportional to the muscle mass. Creatinine is mostly
cleared from the blood by simple glomerular filtration but it is also
Abbreviations: LC, Liquid chromatography; GC, Gas chromatography; MS, Mass
spectrometry; MS/MS, Tandem mass spectrometry; IDMS, Isotope dilution mass
spectrometry; HPLC-UV, High pressure liquid chromatography with ultraviolet
absorbance detection; CKD, Chronic kidney disease; eGFR, Estimated glomerular
filtration rate; SRM, Standard reference material.
⁎ Corresponding author. Biochimie, CHUM, Hôpital Notre-Dame, 1560 Sherbrooke
Est, Montréal, Québec, Canada H2L 4M1. Fax: +1 514 412 7553.
E-mail address: [email protected] (P.-O. Hétu).
slightly secreted by the renal tubules, and this secretion can increase in
compensation of reduced GFR [5]. Thus creatinine is not a perfect
marker, but its clearance is nevertheless routinely used to evaluate GFR.
However, measurement of the rate of creatinine clearance is laborious
and error prone, mostly because of the necessity to properly collect
urine during a 24 h period. In addition, creatinine clearance is usually
only prescribed when the physician suspects a decreased kidney
function. In an effort to better identify patients with CKD in early stages,
it is now recommended to calculate and report an estimated GFR (eGFR)
with every serum creatinine test result generated in the laboratory [6].
An eGFR reported for every patient, irrespectively of clinical signs or
prior suspicion of decreased kidney function, should allow an earlier
referral of CKD patients for proper follow up and potentially retard the
progression of the disease.
Multiple formulas exist for calculating eGFR from serum creatinine, including the Cockcroft–Gault [7], the MDRD (Modification of
Diet in Renal Disease) [8,9], and the CKD-EPI (Chronic Kidney Disease
Epidemiology Collaboration) [10] formulas. However, the use of any
of the formulas to estimate GFR requires exact and precise creatinine
results. For example, a 7.6% under-estimation of the creatinine value
will lead to an overestimation of eGFR of 9.6% when using the MDRD
formula. It has thus been recommended that creatinine results used
for calculating eGFR with the MDRD formula should be traceable to an
isotope dilution mass spectrometric (IDMS) reference method, with
desirable values for bias, imprecision, and total error of respectively
0009-9120/$ – see front matter © 2010 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.
doi:10.1016/j.clinbiochem.2010.05.018
P.-O. Hétu et al. / Clinical Biochemistry 43 (2010) 1158–1162
3.4%, 2.2%, and 7.6% at approximately 88.4 μmol/L creatinine [6]. In
line with this, most in-vitro diagnostic manufacturers have now
calibrated their creatinine assays against an IDMS traceable method.
Definitive methods for quantification of creatinine by gas chromatography coupled to isotope dilution mass spectrometry (GC-IDMS)
have been described and they serve to assign creatinine values to the
standard reference materials [11–14]. GC-IDMS methods have been
certified by the Joint Committee for Traceability in Laboratory
Medicine (JCTLM), an international committee in which participate
the BIPM (Bureau International des Poids et Mesures), the IFCC
(International Federation for Clinical Chemistry and Laboratory
Medicine), and the ILAC (International Laboratory Accreditation
Cooperation). However, the sample pretreatment prior to GC-IDMS
analysis of creatinine is a very laborious multi-step procedure that
includes precipitation of serum proteins, isolation of creatinine by
liquid chromatography, evaporation of the creatinine fraction, and
formation of creatinine derivatives. This complexity leads to higher
costs per sample and to a lower throughput, which limits the
applicability of definitive methods for large-scale validation of routine
laboratory creatinine assays. Unfortunately, too few laboratories in the
world offer the service of a definitive method for serum creatinine by
GC-IDMS, which complicates the standardization of results. More
recently, methods using liquid chromatography (LC-IDMS) have been
developed [14–17], and such a method was used by the National
Institute of Standards and Technology (NIST) to certify the creatinine
concentrations in the two levels of their Standard Reference Material
967 [14].
Here we report the development and validation of a new simple
liquid chromatography isotope dilution tandem mass spectrometry
(LC-IDMS/MS) method for creatinine quantification. The method is
shown to be exact and precise, it is not affected by ion suppression
effects, and it is free of sample carryover. In addition, the sample
preparation is not complex and the method has a high sample
throughput. These characteristics make the method a good candidate
reference method that could be used to validate the IDMS traceability
of routine clinical laboratory methods, and to reduce the differences
that may subsist between different manufacturers' assays.
Methods
Materials
Standard reference materials (SRM) of purified creatinine (SRM
914a) and creatinine in frozen human serum (SRM 967) were purchased
from the NIST (National Institute of Standards and Technology,
Gaithersburg, MD, USA). Deuterated creatinine (creatinine-D3, cat.
#485446) and HPLC grade ammonium acetate, formic acid, and
methanol were purchased from Sigma-Aldrich (Saint-Louis, MO, USA).
1159
Sample processing
Samples were prepared for analysis by transferring 15 μL of
calibrator, control, or serum (using a 25 μL positive displacement
pipette) to a 1.5 mL microcentrifuge tube containing 15 μL internal
standard (10 μg/mL creatinine-D3 in water). Samples were then vortex
mixed briefly and 1.2 mL of methanol was then added to precipitate the
proteins. Samples were briefly vortex mixed, and after 10 min of
mechanical stirring the samples were centrifuged at 11,000 × g for 5 min.
Supernatants were transferred to vials and analyzed by LC-MS/MS.
LC-MS/MS analysis
Sample supernatants (5 μL) were injected on a 1200 series HPLC
system coupled to a 6410 tandem mass spectrometer (Agilent
Technologies). Normal phase chromatography of creatinine was
performed on a strong cation exchanger SecurityGuard pre-column
(SCX, 4 × 3 mm, 10 μm, cat. #AJ0-4308, Phenomenex, Torrance, CA,
USA) at a temperature of 30 °C. Normal phase separation was
achieved in isocratic mode at a flow rate of 0.6 mL/min using a
mobile phase of 75% methanol and 25% of 10 mmol/L ammonium
acetate containing 0.4% formic acid (v/v), in a total chromatographic
run time of 3.0 min (with the first 0.3 min diverted directly to waste
instead of the MS/MS detector). Detection of creatinine and
creatinine-D3 was accomplished by tandem mass spectrometry in
positive mode using multiple reaction monitoring with ion transitions
of 114→44 and 117→47 m/z, respectively. Details of the mass
spectrometry settings are listed in Table 1. The mass spectrometer
was calibrated with 5 creatinine levels in duplicate, once at the
beginning of the batch and once again at the end. The difference
between duplicate curves was less than 1%, so the average of both
curves was used for quantification of the controls and patient samples.
Accuracy and precision
Intra-assay precision was determined by quantifying creatinine in
both levels of the standard reference material SRM 967 (5 times each),
intercalated in a single series of 45 patient samples, while the values
from 6 independent analytical series performed over 2 months were
used to calculate the inter-assay precision. The accuracy was
evaluated by comparing the average concentration of creatinine
measured for both levels of SRM 967 with the NIST certified target
values of 66.5 and 346.2 μmol/L, respectively [14]. In addition, the
accuracy of the LC-IDMS/MS method was further validated by adding
10 μL of purified creatinine (8.95 mmol/L in 0.005 mol/L HCl) to
500 μL of different patient serums (n = 10, range of 31–338 μmol/L
creatinine). The measured increase in creatinine concentration was
compared to the expected increase of 175.5 μmol/L.
Quantification of sample carryover
Preparation of calibrators
Creatinine standards were prepared gravimetrically. To prepare the
1 mg/mL primary standard, approximately 100 mg of SRM 914a was
precisely weighed and transferred to a pre-weighed volumetric flask.
The creatinine was then dissolved in 100 mL of HCl (0.005 mol/L), and
the flask was weighed again. The concentration of creatinine in the
primary standard was obtained by calculating the weight of creatinine
using the certified purity of SRM 914a (99.8%) and the exact volume of
water using a density of 0.997 g/mL (at 25 °C). The calculated
concentration was converted to SI units (μg/mL× 8.84 = μmol/L). The
highest calibrator for the calibration curve was then prepared
gravimetrically by diluting the primary creatinine standard with
water, and the other calibrators were obtained by serial dilutions with
water using a recently calibrated positive displacement pipette.
Sample carryover was evaluated according to the method of
Broughton [18]. Briefly, two patient samples, one with high (H) and
Table 1
Mass spectrometer parameters.
Source parameters
Gas temp: 350 °C
Gas flow: 10 L/min
Nebulizer: 50 psi
Capillary voltage: 1200 V
Acquisition parameters
Fragments (m/z):
Creatinine: 114→44
Creatinine-D3: 117→47
Fragmentor voltage: 100 V
Dwell time: 50 ms
Collision energy: 20 V
1160
P.-O. Hétu et al. / Clinical Biochemistry 43 (2010) 1158–1162
the other with low (L) creatinine concentration (837 μmol/L and
32 μmol/L, respectively), were sequentially analyzed in triplicate
(H1, H2, H3 followed by L1, L2, L3). The percentage of sample
carryover was calculated by the formula (L1 − L3) ÷ (H3 − L3) × 100,
and the average of 3 such series was defined as the method's sample
carryover.
Evaluation of ion suppression
The presence of ion suppression effects on the MS/MS response
was evaluated by spiking 110.6 μmol/L creatinine in a pool of serum
or in water (n = 5). The creatinine peak area corresponding to the
added creatinine in the pool of serum was calculated by subtracting
the endogenous creatinine from the total creatinine measured in the
spiked serums. The MS/MS response, in absolute counts of peak area,
calculated for the added creatinine in serum was then compared to
that measured in water.
Correlation with an automated method
Results for patient samples analyzed by the LC-IDMS/MS method
were compared to the results obtained from an IDMS traceable kinetic
Jaffe colorimetric assay on an automated analyzer (Hitachi 917, Roche
Diagnostics). All samples were analyzed in duplicate (n = 36) and the
correlation between the methods was computed by Deming linear
regression analysis using the CBstat version 5.1.1 software (K. Linnet,
purchased from the AACC online store, cat. #2475, www.aacc.org).
Table 2
Method accuracy (mean ± SD).
Sample
Target value
(μmol/L)
Measured value
(μmol/L)
% Recovery
(measured/target)
SRM 967, level 1
(n = 5)
SRM 967, level 2
(n = 5)
Spiked samples
(n = 10)
66.5 (62.7–70.3)
64.6 ± 0.6
97.1 ± 0.9
346.2 (331.6–360.8)
354.5 ± 3.2
102.1 ± 0.9
175.5
181.8 ± 7.2
103.6 ± 4.1
Results
A cost-effective, rapid, and simple LC-IDMS/MS method for the
quantification of serum creatinine was developed. Even if a cation
exchange chromatography material was used, the principal mechanism of
separation is believed to be normal phase partition because the retention
time of creatinine increased with increasing proportions of organic
modifier in the mobile phase (data not shown). As expected, the higher
methanol content in the mobile phase also resulted in a higher sensitivity
on the mass spectrometer. In the chosen conditions, both creatinine and
the deuterated internal standard eluted at approximately 1.6 min (Fig. 1A)
in a total chromatographic time of 3 min. An injection overlap was also
programmed, resulting in a throughput of approximately 17 samples per
hour. Calibration of each analytical run was done in duplicate, at the start
and at the end of the batch, and no drift in MS/MS relative response of
creatinine/creatinine-D3 was observed between the duplicates (Fig. 1B).
The calibrators showed a linear response on the mass spectrometer (from
34.6 to 553.3 μmol/L creatinine), although the data were better fitted by a
second order polynomial function, as seen regularly in LC-MS.
Accuracy and precision
The proposed method was shown to be accurate by analyzing the
standard reference material from the NIST (SRM 967, levels 1 and 2) and
by spiking known amounts of creatinine in 10 patient samples (Table 2).
The concentrations of creatinine measured for both levels of SRM 967
were accurate (97.1–102.1% of target value), and the added creatinine was
completely recovered in the spiked patient samples. A higher variability
was observed for the spiking experiments, partially because only a small
volume of creatinine was added to the patient samples (to lessen the
impact on the sample matrix), and also because the calculated recovery
includes the imprecision of the pre- and post-spiking creatinine
concentrations. The method was also very precise, with intra-assay
precision below 1% and inter-assay precision below 2% for both levels of
the NIST reference material (Table 3).
Since creatinine-free serum samples were not available, the
analytical sensitivity was determined using a sample with approximately 65 μmol/L creatinine (SRM 967, level 1), which is at the low
end of the practical analytical range (i.e., the low end of the reference
interval). The calculated analytical sensitivity, defined as three times
the standard deviation observed for the SRM 967 level 1, was
1.7 μmol/L. The creatinine concentration of the lowest standard on
the calibration curve (34.6 μmol/L) was defined as the lower limit of
quantification, in accordance with published recommendations for
chromatographic methods [19].
Table 3
Method precision.
Sample
Fig. 1. Typical chromatograph and calibration curve. A) Chromatograph of the NIST SRM
967 (level 1). The MS/MS response for creatinine (solid line, shaded peak) is overlaid
on the response of the creatinine-D3 internal standard (dotted line). B) The 5 point
calibration curve fitted with a second order polynomial equation.
Intra-assay precision (n = 5)
SRM 967, level 1
SRM 967, level 2
Inter-assay precision (n = 6)
SRM 967, level 1
SRM 967, level 2
Mean (μmol/L)
CV
64.6
354.5
0.89%
0.89%
64.7
353.3
1.9%
1.8%
P.-O. Hétu et al. / Clinical Biochemistry 43 (2010) 1158–1162
Table 4
Ion suppression (n = 5).
Sample
Serum
Water
Endogenous creatinine
Total creatinine (after spiking)
Added creatinine (spiked)
Added creatinine (spiked)
Peak area (counts)
CV
3.49 × 105
1.09 × 106
7.40 × 105
7.29 × 105
0.93%
1.58%
2.76%
1.75%
Creatinine (110.6 μmol/L) was spiked in a pool of serum or in water. The peak area
corresponding to the added creatinine in serum was calculated by subtracting the
endogenous creatinine from the total creatinine concentration after spiking.
Sample carryover and matrix effects
The method was shown to be free of significant sample carryover
by multiple sequential analyses of three high concentration samples
followed by three low concentration samples (0.14% carryover, n = 3).
The absence of an ion suppression effect on the MS/MS response was
confirmed by spiking creatinine in a pool of serum or in water (Table 4).
There was no significant difference between the spiked creatinine peak
areas in water or serum (serum/water= 101.6 ± 2.9%, p = n.s.).
Correlation with an automated method
Finally, a routine IDMS traceable kinetic Jaffe method on the Roche
Modular analyzer was compared with the LC-IDMS/MS method by
analyzing 36 patient samples in duplicate with both methods (range
of samples from 26 to 575 μmol/L creatinine). Deming regression
Fig. 2. Correlation between LC-IDMS/MS and an automated kinetic Jaffe assay. A) The
creatinine concentration in patient samples (n = 36) was determined in duplicate with
both methods, and the results obtained with the automated assay were compared to
those obtained by LC-IDMS/MS. Deming regression analysis results are presented, with
95% confidence interval of the slope depicted as dotted lines. B) Bland–Altman plot
showing the relative difference between both methods, with the mean relative bias and
the corresponding 95% confidence interval represented by dotted lines.
1161
analysis showed that the creatinine results obtained by the colorimetric assay correlated very well with the LC-IDMS/MS method
(Fig. 2A), although a small but statistically significant constant bias of
approximately 3.4 μmol/L was observed (p b 0.002). The calculated
mean relative bias was 4.3% (Fig. 2B), and all but two data points fell
within the 95% confidence interval (−5.8 to 14.5%).
Discussion
We present here the development and validation of a candidate
reference IDMS method for the quantification of creatinine. The
method is highly sensitive, it is exact and precise, it is free of
significant sample carryover and it is not affected by ion suppression
effects. Together with the relatively high throughput, achieved by the
simple sample preparation and the sole use of a short cartridge for
chromatography, these characteristics make the method ideal for the
validation of the IDMS traceability of routine assays.
Other mass spectrometric methods using deuterated creatinine as
internal standard (IDMS methods) have been published [11–17].
Although the certified definitive methods are GC-IDMS methods [11–
13], LC-IDMS methods were recently described and have been used
for assigning certified values to reference materials [14]. The LC-IDMS
methods have the advantage of being much less labor intensive than
GC-IDMS methods because derivatization of creatinine is required
prior to GC analysis, which is not the case for LC-IDMS. A preliminary
separation of creatinine from creatine by liquid chromatography must
also be performed before GC analysis, since these two compounds
become undistinguishable once derivatized [14,20]. The development
of liquid chromatography based reference IDMS methods should thus
be encouraged, in part because of the higher complexity and the lower
throughput of GC-IDMS definitive methods compared to LC-IDMS, but
mostly because results from both methodologies have been shown to
be comparable [14,15].
Different strategies for the chromatography of creatinine can be
employed in LC methods. Most previously published methods have
used reversed phase [14–16] or ion exchange [17,21,22] chromatography. For our method, we chose to use a small cation exchange precolumn, as described previously by Owen et al. [17]. However, the
method of Owen et al. was not calibrated nor validated using certified
materials, and the intra- and inter-assay imprecisions were higher
than those observed with the present method. Also, instead of the ion
exchange chromatography proposed by Owen et al., normal phase
partition chromatography retention of creatinine on the SCX cartridge
was favored in our method by using a high proportion of methanol in
the mobile phase. This strategy resulted in good retention of creatinine
on the cartridge and was associated with very good sensitivity because
of the higher volatility of the mobile phase that yields better ionization
efficiency on the mass spectrometer. Additionally, it allowed the
presence of higher proportions of methanol in the injected sample
without affecting the chromatographic performance. The simple
protein precipitation used for the pretreatment step was sufficient
for sample clean-up, as previously published [15]. Thus, the proposed
LC-IDMS/MS method for serum creatinine quantification is simple,
accurate, precise, and sensitive, with a high throughput and a
relatively low cost. The described method is proposed as a candidate
reference method, because of its low imprecision (b2%) and its
accuracy in the quantification of the certified reference material SRM
967. Unfortunately, it was not possible to compare our method with a
definitive method before writing this paper because of the limited
worldwide availability of certified GC-IDMS methods.
The laboratory plays a significant role in the proper reporting of
eGFR by assuring the production of reliable serum creatinine results,
and this is capital for the correct identification and staging of CKD
patients. In addition, the comparability of creatinine test results
produced by different laboratories within a defined geographical
region is important for the long term follow up of CKD patients, who
1162
P.-O. Hétu et al. / Clinical Biochemistry 43 (2010) 1158–1162
are likely to request laboratory tests from different hospitals.
Although most manufacturers, if not all, now offer IDMS traceable
creatinine assays, differences can still subsist between assays.
Different hospitals have different creatinine assays from different
manufacturers, so the comparability of the results between laboratories must be assessed, and eventually corrected, by external quality
control and educational programs. For example, the successful
implementation of a provincial creatinine standardization program
in British Columbia (Canada), before the IDMS standardization of
assays by instrument manufacturers, has been described [23]. This
program, directed by the Canadian External Quality Assessment
Laboratory (CEQUAL), consisted in sending GC-IDMS certified serum
aliquots to all participating laboratories and by attributing proper
correction factors to normalize the results of creatinine assays. With
this strategy, the average bias and the average total error for the
measurement of creatinine were reduced from 16.5 to 2.7% and from
23.9 to 8.7%, respectively [23].
The harmonization of creatinine results within a defined region,
for example a province or a state, should be encouraged and the tools
to achieve this goal need to be made available. To this end, the method
presented here was developed to allow laboratories in the province of
Quebec (Canada) to validate the IDMS traceability of their routine
method, on a voluntary basis. The relatively low cost of the proposed
LC-IDMS/MS reference method and the higher throughput compared
to the definitive GC-IDMS methods should encourage the participation of laboratories to the provincial program, leading to more
standardized creatinine results between different hospitals, to better
estimation of GFR, and to better patient care.
Acknowledgments
The authors would like to thank, the SQBC (Société québécoise de
biologie clinique) for financial support. The authors are also grateful to
Dr M. Bouthiller (Hôpital de Granby, QC, Canada) for quantifying
creatinine in patient samples with the Hitachi 917 analyzer and for
providing these samples to allow comparison with our LC-IDMS/MS
method.
References
[1] Coresh J, Selvin E, Stevens LA, et al. Prevalence of chronic kidney disease in the
United States. JAMA 2007;298:2038–47.
[2] Levin A, Hemmelgarn B, Culleton B, et al. Guidelines for the management of
chronic kidney disease. CMAJ 2008;179:1154–62.
[3] Anavekar NS, McMurray JJ, Velazquez EJ, et al. Relation between renal dysfunction
and cardiovascular outcomes after myocardial infarction. N Engl J Med 2004;351:
1285–95.
[4] Hillege HL, Nitsch D, Pfeffer MA, et al. Renal function as a predictor of outcome in a
broad spectrum of patients with heart failure. Circulation 2006;113:671–8.
[5] Schwartz GJ, Furth SL. Glomerular filtration rate measurement and estimation in
chronic kidney disease. Pediatr Nephrol 2007;22:1839–48.
[6] Myers GL, Miller WG, Coresh J, et al. Recommendations for improving serum
creatinine measurement: a report from the Laboratory Working Group of the
National Kidney Disease Education Program. Clin Chem 2006;52:5–18.
[7] Cockcroft DW, Gault MH. Prediction of creatinine clearance from serum creatinine.
Nephron 1976;16:31–41.
[8] Levey AS, Bosch JP, Lewis JB, et al. A more accurate method to estimate glomerular
filtration rate from serum creatinine: a new prediction equation. Modification of
Diet in Renal Disease Study Group. Ann Intern Med 1999;130:461–70.
[9] Levey AS, Coresh J, Greene T, et al. Expressing the Modification of Diet in Renal
Disease Study equation for estimating glomerular filtration rate with standardized
serum creatinine values. Clin Chem 2007;53:766–76672.
[10] Levey AS, Stevens LA, Schmid CH, et al. A new equation to estimate glomerular
filtration rate. Ann Intern Med 2009;150:604–12.
[11] Siekmann L. Determination of creatinine in human serum by isotope dilutionmass spectrometry. Definitive methods in clinical chemistry, IV. J Clin Chem Clin
Biochem 1985;23:137–44.
[12] Welch MJ, Cohen A, Hertz HS, et al. Determination of serum creatinine by isotope
dilution mass spectrometry as a candidate definitive method. Anal Chem 1986;58:
1681–5.
[13] Stockl D, Reinauer H. Candidate reference methods for determining target values
for cholesterol, creatinine, uric acid, and glucose in external quality assessment
and internal accuracy control. I. Method setup. Clin Chem 1993;39:993–1000.
[14] Dodder NG, Tai SS, Sniegoski LT, Zhang NF, Welch MJ. Certification of creatinine in
a human serum reference material by GC-MS and LC-MS. Clin Chem 2007;53:
1694–9.
[15] Stokes P, O'Connor G. Development of a liquid chromatography-mass spectrometry method for the high-accuracy determination of creatinine in serum. J
Chromatogr B Analyt Technol Biomed Life Sci 2003;794:125–36.
[16] Takahashi N, Boysen G, Li F, Li Y, Swenberg JA. Tandem mass spectrometry
measurements of creatinine in mouse plasma and urine for determining
glomerular filtration rate. Kidney Int 2007;71:266–71.
[17] Owen LJ, Wear JE, Keevil BG. Validation of a liquid chromatography tandem mass
spectrometry assay for serum creatinine and comparison with enzymatic and Jaffe
methods. Ann Clin Biochem 2006;43:118–23.
[18] Broughton PMG, Buttolph MA, Gowenlock A, Neill DW, Skentelbery RG.
Recommended scheme for the evaluation of instruments for automatic analysis
in the clinical biochemistry laboratory. J Clin Pathol 1969;22:278–84.
[19] US Department of Health and Human Services. Guidance for industry: bioanalytical method validation. Food and Drug Administration, Center for Drug Evaluation
and Research (CDER), Center for Veterinary Medicine (CVM). 2001.
[20] Peake M, Whiting M. Measurement of serum creatinine - current status and future
goals. Clin Biochem Rev 2006;27:173–84.
[21] Rosano TG, Ambrose RT, Wu AH, Swift TA, Yadegari P. Candidate reference method
for determining creatinine in serum: method development and interlaboratory
validation. Clin Chem 1990;36:1951–5.
[22] Ambrose RT, Ketchum DF, Smith JW. Creatinine determined by “high-performance” liquid chromatography. Clin Chem 1983;29:256–9.
[23] Komenda P, Beaulieu M, Seccombe D, Levin A. Regional implementation of
creatinine measurement standardization. J Am Soc Nephrol 2008;19:164–9.
Téléchargement