absorption cell should not contain metallic parts that contact the
eluent-sample flow. The cell must be usable at 530 nm. Use
plastic pressurized containers to deliver eluent and post-column
reagent. Use high-purity helium (99.995%) to pressurize the
eluent and post-column reagent vessel.
b. Guard column, to be placed before the separator column,
containing an adsorbent capable of adsorbing organic com-
pounds and particulates that would damage or interfere with the
analysis or equipment.*
c. Separator column, packed with a high-capacity anion-ex-
change resin capable of resolving chromate from other sample
constituents.†
d. Recorder, integrator, or computer for receiving signals
from the detector as a function of time.
e. Laboratory ware: Soak all reusable items (glass, plastic,
etc.) including sample containers, overnight in laboratory-grade
detergent, rinse, and soak for4hinamixture of nitric acid (1
part), hydrochloric acid (2 parts), and reagent water (9 parts).
Rinse with tap water and reagent water. NOTE: Never use chro-
mic acid cleaning solution.
f. Syringe, equipped with male luer-type fitting and a capacity
of at least 3 mL.
g. pH meter: Use any commercial pH meter. Standardize and
calibrate according to manufacturer’s instructions. Pay special
attention to temperature compensation and electrode care.
3.
Reagents
a. Reagent water: Deionized or distilled water free from
interferences at the minimum detection level of each constituent,
filtered through a 0.2-
m membrane filter and having a conduc-
tance of less than 0.1
S/cm. Use for preparing all reagents (see
Section 1080).
b. Cr(VI) stock solution, 100 mg Cr
6⫹
/L: Prepare from pri-
mary standard grade potassium dichromate. Dissolve 0.1414 g
K
2
Cr
2
O
7
in water and dilute to 500 mL in a volumetric flask. pH
adjustment is not required. Store in plastic. CAUTION: Hexava-
lent chromium is toxic and a suspected carcinogen; handle
with care.
c. Eluent: Dissolve 33 g ammonium sulfate (NH
4
)
2
SO
4
,in500
mL water and add 6.5 mL conc ammonium hydroxide, NH
4
OH.
Dilute to 1 L with water.
d. Post-column reagent: Dissolve 0.5 g 1,5-diphenylcarbazide
in 100 mL HPLC-grade methanol. Add with stirring to 500 mL
water containing 28 mL conc H
2
SO
4
. Dilute to 1 L with water.
Reagent is stable for 4 or 5 d; prepare only as needed.
e. Buffer solution: Dissolve 33 g ammonium sulfate,
(NH
4
)
2
SO
4
, in 75 mL water and add 6.5 mL conc ammonium
hydroxide, NH
4
OH. Dilute to 100 mL with water.
f. Sodium hydroxide, 5N: Dissolve 200 g NaOH in 1 L water.
Store in plastic bottle. Smaller volumes may be prepared.
4.
Procedure
a. Instrument setup: Establish ion chromatograph operating
conditions as indicated in Table 3500-Cr:I. Set flow rate of the
eluent pump at 1.5 mL/min and adjust pressure of reagent
delivery module so that the system flow rate, measured after the
detector, is 2.0 mL/min. Measure system flow rate using a
graduated cylinder and stopwatch. Allow approximately 30 min
after adjustment before measuring flow.
Use an injection loop size based on required sensitivity. A
50-
L loop is sufficient, although a 250-
L loop was used to
determine the method detection level.
b. Calibration: Before sample analysis, construct a calibration
curve using a minimum of a blank and three standards that
bracket the expected sample concentration range. Prepare cali-
bration standards from the stock standard (3b) by appropriate
dilution with reagent water in volumetric flasks. Adjust to pH 9
to 9.5 with buffer solution (3e) before final dilution. Injection
volumes of standards should be about 10 times the injection loop
volume to insure complete loop flushing.
c. Sample analysis: Bring chilled, pH-adjusted sample to
ambient temperature. Fill a clean syringe with sample, attach a
0.45-
m syringe filter, and inject 10 times the sample loop
volume into the instrument. Dilute any sample that has a con-
centration greater than the highest calibration standard.
5.
Calculation
Determine area or height of the Cr(VI) peak in the calibration
standard chromatograms. Establish a calibration curve by per-
forming a regression analysis of peak height or peak area against
standard concentration in mg/L. If correlation coefficient is less
than 0.995, do not use these data. Check analytical process.
For samples, measure area (or height) of Cr(VI) peak in
sample chromatogram, as determined by retention time. Calcu-
late Cr(VI) concentration by interpolating from the calibration
line. Correct data for any dilutions made.
Currently available instrumentation automates the entire mea-
surement process (peak measurement, calibration, and sample mea-
surements and calculations). Ensure that enough quality control
samples are analyzed to monitor the instrumental processes.
6.
Quality Control
The quality control practices considered to be an integral part
of each method can be found in Section 3020.
a. Initial demonstration of performance: Before sample anal-
ysis, set up instrument and analyze enough known samples to
* IonPac NG1, Dionex, 4700 Lakeside Drive, Sunnyvale, CA 94086, or equivalent.
† IonPac AS7, Dionex, or equivalent.
TABLE 3500-Cr:I. ION CHROMATOGRAPHIC CONDITIONS
Variable Value
Guard column Dionex IonPac NG1
Separator column Dionex IonPac AS7
Eluent 250 mM(NH4)
2
SO
4
100 mMNH
4
OH
Eluent flow rate 1.5 mL/min
Post-column reagent 2 mMdiphenylcarbohydrazide
10% v/v CH
3
OH
1NH
2
SO
4
Post-column reagent flow rate 0.5 mL/min
Detector Visible @ 530 nm
Retention time 3.8 min
CHROMIUM (3500-Cr)/Ion Chromatographic Method
4
CHROMIUM (3500-Cr)/Ion Chromatographic Method