If using response factors or calibration factors, relative stan-
dard deviation (RSD) for each compound of interest should be
less than 20%. If the RSD is not less than 20% for any compound
of interest, then identify and correct source of lack of linearity
before sample quantitation. When using response factors (i.e.,
for GC/MS analysis), check performance or sensitivity of the
instrument for the compound of interest against minimum ac-
ceptance values for the response factors. See specific analytical
method for the acceptance criteria for the response factors for
each compound.
For a linear regression, the correlation coefficient should be
⬎0.994. Recalculate each calibration point compared to curve.
Resulting values should be within ⫾20%. If any of the recalcu-
lated values are not within ⫾20%, identify and correct source of
outlier(s) before sample quantitation.
Use initial calibration, with any of the above functions (re-
sponse factor, calibration factor, or calibration curve) for quan-
titation of the analytes of interest in samples. Use continuing
calibration, described in ¶ 2) below, only for checks on initial
calibration and not for sample quantitation. Perform initial cal-
ibration when instrument is set up and whenever continuing
calibration criteria are not met.
2) Continuing calibration—Continuing calibration (CCAL)
is the periodic analysis of a calibration standard used to verify
that the instrument response has not changed significantly
from the initial calibration. Perform continuing calibration
every 10 samples for GC analysis, every 20 samples for
GC/MS analysis, or every 12 h, whichever is more frequent.
Perform continuing calibration with one or more of the con-
centrations of analytical standards in the initial calibration.
Vary actual concentration of continuing calibration standard
over calibration range, with a minimum concentration greater
than two times the reporting limit. The acceptance criterion
for continuing calibration is 70 to 130% recovery compared to
the known or expected value of the calibration standard (at the
analyst’s discretion, the acceptance criterion for the gases
may be extended to 60 to 140% recovery). If the acceptance
criteria are not met, re-analyze continuing calibration stan-
dard or repeat initial calibration. When using response fac-
tors, check performance or sensitivity of instrument for ana-
lytes of interest against minimum acceptance values for re-
sponse factors.
3) Closing standard—Finish all sample sets with a closing stan-
dard to demonstrate that performance was still acceptable for the
last sample analyzed. Use acceptance criteria as for the CCAL.
c. Batch quality control:
1) Analytical day—An analytical day is defined as a 12-h
analytical period.
2) Sample set (batch)—A sample set (batch) is defined as
those samples extracted in an analytical day, not to exceed 20
samples.
3) Laboratory reagent blank (LRB)—A LRB is a blank sam-
ple consisting of all reagents that normally contact a sample
when carried through the entire analytical procedure. Use re-
agent blank to determine contribution of reagents and prepara-
tive analytical steps to observed value. No compound of interest
should be present in reagent blank at a level greater than the
MQL. Include a minimum of one reagent blank with each sample
set (batch).
4) Laboratory-fortified blank (LFB)—See ¶ b2) above. NOTE:
For this method, the LFB and CCAL are the same.
5) Internal standard (IS)—An internal standard is a compound
of known concentration added to each standard and sample just
before sample analysis. Because of the nature of purge and trap
analysis, the IS is taken through the entire analytical process, just
as is the surrogate standard [see ¶ 6) below]. However, the IS is
used for quantitation, whereas the surrogate standard is used to
monitor ongoing purge recovery. Use IS to monitor retention
time, relative response, and concentration of analytes in each
sample. When quantifying by the internal standard method,
measure all compound responses relative to this standard. Inter-
nal standard area response should be in the range of ⫾30%
compared to the mean calibration curve area response. The IS
compound should mimic the chromatographic conditions of the
analytes of interest. The retention time of this compound should
separate from all analytes of interest and elute in a representative
area of the chromatogram. If a specific compound cannot be
found to meet these criteria, use additional compounds to satisfy
analytical needs.
6) Surrogate standard (SS)—A surrogate standard is a com-
pound added to each standard and sample at a known concen-
tration before extraction. Choose a compound(s) that is chemi-
cally similar to the analytes and that is unlikely to be found in
environmental samples. Carry surrogate standard through entire
sample extraction and analytical process to monitor extraction
recovery for each sample. Surrogate recovery should remain
reasonably constant over time. Recovery should not vary more
than 30% from the known value. Refer to method of choice for
specific surrogates.
7) Laboratory-fortified sample (LFS)—A LFS is an additional
portion of a sample to which the analytes of interest have been
added at a concentration at least two times the MRL or around
the middle of the calibration range. Include a minimum of one
LFS with each sample set (batch). Make LFSs at sufficient
concentrations that sample background levels do not adversely
affect recovery calculations. (If this is a known sample, adjust
addition concentrations to be about five times background level).
Base sample batch acceptance on results of CCALs and LFBs
rather than on LFSs, because the matrix of the sample may
interfere with method performance. Prepare fortifying solution
for blanks and samples from a different primary mix than that
used to develop working standard mix.
8) LFS duplicates—A LFS duplicate is a second LFS used to
evaluate the precision of the method in a matrix sample. If
sufficient sample volume is collected, fortify a large enough
volume to yield two sample portions for analysis. If sufficient
sample volume is not collected, use a second bottle of the same
sample fortified to the same concentration as the first. Include a
minimum of one LFS duplicate with each sample set (batch).
Compare precision and bias to those listed in the method. Base
sample batch acceptance on results of CCAL and LFB additions
rather than LFS duplicates.
6. References
1. U.S. ENVIRONMENTAL PROTECTION AGENCY. 1987. National Primary
Drinking Water Regulations—synthetic organic chemicals; monitor-
ing for unregulated contaminants; final rule. 40 CFR 141 & 142;
Federal Register 52, No. 130.
VOLATILE ORGANIC COMPOUNDS (6200)/Introduction
2
VOLATILE ORGANIC COMPOUNDS (6200)/Introduction