1 h. Clean caps and septa by rinsing with acetone, then hexane.
Heat at 80°C for not more than1hinaclean, forced-air
convection oven.
b. Microsyringes, 5, 10, 25, 50, 100, 250, 500, and 1000
L.
c. Syringe, 30-mL glass hypodermic, metal luer-lok tip with
8.9-cm- (3.5-in.-) long ⫻17 gauge stainless steel pipetting
needle (alternatively use a 30-mL volumetric pipet). See ¶ a
above for glassware cleaning procedure.
d. Micro volumetric flasks,† TFE-lined screw-cap: 2-mL,
5-mL, and 10-mL. Immediately after use, rinse three times with
methanol. Invert to drain. Let air-dry completely in a ventilation
hood.
e. Mechanical shaker,‡ to automate MtBE extraction. Insert
vials into a wooden holding block (20 vial capacity) made from
laminated plywood with drilled holes to accept vials, dimen-
sioned to fit snugly onto the shaker table.
f. Extract and standard solution storage container, 1.8-mL
clear glass, 7- and 14-mL amber glass screw-cap vials with
TFE-lined silicone septa. For cleaning procedures, see ¶ aabove.
g. Transfer pipets, 14.6- and 23-cm (5.75- and 9-in.) dispos-
able glass pasteur pipets. See ¶ aabove for glassware cleaning
procedure.
h. Gas chromatograph, temperature-programmable (prefera-
bly with multiple ramp capability) with injector. Optimally use
an autosampler for sample injection and a computer data system
for peak integration and quantitation. (A detector base that can
mount two electron capture detectors is ideal.)
1) Gas handling equipment: Use carrier (helium) and makeup
(nitrogen) gases of high purity (99.999%) grade that pass
through indicating calcium sulfate, molecular sieve 5A, activated
charcoal, and an oxygen purifying cartridge. Use two-stage
metal diaphragm high-purity regulators at the compressed gas
sources. Use flow controllers to regulate carrier gas flow. Make
all gas lines 0.3-cm (1/8-in.) copper (or stainless steel) tubing;
rinse with acetone and bake before use.
2) Injector, split/splitless (using straight open bore insert).
3) Analytical column,§ 30 m long ⫻0.25 mm ID, fused silica
capillary column with a 0.25-
m film thickness or equivalent.
4) Confirmation column, 30 m long ⫻0.25 mm ID, fused
silica capillary column㛳with a 0.25-
m film thickness or a 30 m
long ⫻0.25 mm ID, fused silica capillary column# with a
0.5-
m film thickness.
5) Detectors, a constant-current pulse-modulated
63
Ni ECD
with standard size cell (use two ECDs for simultaneous confir-
mation analysis).
i. Salt scoops for sodium sulfate, made from stainless steel
1.3-cm- (0.5-in.-) diam bar stock drilled out to a volume of
1.73 mL so that each level scoopful contains 3 g. Alternatively,
weigh the salt.
j. Pipetting dispensers, adjustable 5- and 2-mL sizes with TFE
transfer lines, that can be mounted on the supplier’s reagent
bottles. Use for dispensing H
2
SO
4
and MtBE. Alternatively use
a 3-mL volumetric pipet and a 5-mL graduated pipet with
manual pipet bulbs.
k. Diazomethane generator: Use millimole-size generator
with “o”-ring joint (Figure 6251:2).** Immediately after use,
rinse inner tube twice with 20% NaOH, then rinse twice with tap
water. Immediately add 1 g silica gel to the outside tube to
quench any residual diazomethane solution, rinse twice with
methanol, and twice with tap water. Rinse both inner and outer
tubes with reagent water three times. Bake at 180°C until dry in
a clean, forced-air convection oven.
Alternatively use the millimole-size diazomethane generator
shown in Figure 6251:3.†† To clean, rinse with reagent water
and then with methanol, invert, and let air-dry.
l. pH-strips, pH indicating strips, 0 to 2.5 range.
4.
Reagents
a. Extraction solvent,99⫹% MtBE, preservative-free.‡‡
b. Sodium sulfate reagents:
1) Granular sodium sulfate,Na
2
SO
4
, reagent grade, suitable
for pesticide analysis. Heat at 400°C overnight in a shallow
stainless steel pan covered with aluminum foil. Store in a 1-L
glass bottle with TFE-lined polypropylene cap.
2) Acidified sodium sulfate: To 100 g anhydrous sodium
sulfate, heated as above and cooled, add diethyl ether to just
cover the solid; make a slurry. Add 0.1 mL conc sulfuric acid
and mix thoroughly. Remove ether under low vacuum. Mix 1 g
acidified sodium sulfate with 5 mL reagent water and check that
pH is less than 4. Store remainder of reagent at 130°C.
c. Methanol, pesticide grade or equivalent.
d. Ammonium chloride, NH
4
Cl.
e. Reagents for diazomethane generation with the apparatus
shown in Figure 6251:2:
1) Sodium hydroxide solution, NaOH, 20%: Dissolve 200 g
ACS low-carbonate-grade pellets in 800 mL reagent water.
2) 1-methyl-3-nitro-1-nitrosoguanidine, MNNG.§§
f. Reagents for diazomethane generation with apparatus
shown in Figure 6251:3:
† Kontes, or equivalent.
‡ Eberbach, or equivalent.
§ Durabond-1701, J&W Scientific, or equivalent.
㛳Durabond-5, J&W Scientific, or equivalent.
# Durabond-210, J&W Scientific, or equivalent.
** Aldrich, or equivalent.
†† Paxton Woods Glass, Cincinnati, OH, or equivalent.
‡‡ Omnisolv, manufactured by EM Science, Gibbstown, NJ, or equivalent.
§§ Aldrich or equivalent.
TABLE 6251:I. M
ETHOD
D
ETECTION
L
EVELS AND
P
RECISION
D
ATA*
Compound
Added
Conc.
g/L
Found
Conc.
g/L
Standard
Deviation
g/L
Relative
Standard
Deviation
%
Method
Detection
Level
g/L
Monochloroacetic acid 0.50 0.54 0.026 4.8 0.082
Monobromoacetic acid 0.50 0.80 0.028 3.4 0.087
Dichloroacetic acid 0.50 0.5 0.017 3.5 0.054
Trichloroacetic acid 0.50 0.5 0.017 3.4 0.054
Bromochloroacetic
acid 0.50 0.49 0.015 3.1 0.04
Dibromoacetic acid 0.50 0.47 0.021 4.4 0.065
2,4,6-Trichlorophenol 0.25 0.27 0.011 4.1 0.034
* Based on the analysis of seven portions of reagent water with known additions.
2
DISINFECTION BY-PRODUCTS: HAA (6251)/Micro Liquid-Liquid Extraction GC Method
3
DISINFECTION BY-PRODUCTS: HAA (6251)/Micro Liquid-Liquid Extraction GC Method